Literatur vom gleichen Autor/der gleichen Autor*in
plus bei Google Scholar

Bibliografische Daten exportieren
 

Chemical sensors based on SAW resonators working at up to 1 GHz

Titelangaben

Fischerauer, Gerhard ; Dickert, Franz Ludwig ; Forth, Peter ; Knauer, Ulrich:
Chemical sensors based on SAW resonators working at up to 1 GHz.
In: 1996 IEEE Ultrasonics Symposium Proceedings. - Piscataway, NJ : IEEE , 1996 . - S. 439-442
ISBN 0-7803-3615-1
DOI: https://doi.org/10.1109/ULTSYM.1996.584007

Abstract

We have investigated SAW chemical sensors working at 434 MHz and at 1 GHz, the latter frequency being the highest value reported in this context to date. In particular, the theoretical claim that the gas sensitivity is proportional to the frequency has been checked experimentally. The results of a noise characterization clearly show that the detection limit decreases with frequency even when noise is taken into account. By using molecular cavity compounds such as calixarenes, cyclodextrines and cyclophanes or "molecularly imprinted" polymers as chemically sensitive coatings, sub-ppm traces of halogenated and aromatic hydrocarbons such as toluene or xylene have been detected under ambient conditions.

Weitere Angaben

Publikationsform: Aufsatz in einem Buch
Begutachteter Beitrag: Nein
Keywords: Chemical sensors; hydrocarbons; surface acoustic waves; SAW; sensitivity; frequency dependence
Institutionen der Universität: Fakultäten > Fakultät für Ingenieurwissenschaften > Lehrstuhl Mess- und Regeltechnik > Lehrstuhl Mess- und Regeltechnik - Univ.-Prof. Dr.-Ing. Gerhard Fischerauer
Titel an der UBT entstanden: Nein
Themengebiete aus DDC: 600 Technik, Medizin, angewandte Wissenschaften > 620 Ingenieurwissenschaften
Eingestellt am: 24 Jul 2024 12:01
Letzte Änderung: 24 Jul 2024 12:01
URI: https://eref.uni-bayreuth.de/id/eprint/90035