Spherical sample holders to improve the susceptibility measurement of superparamagnetic materials

Titelangaben

Friedrich, Thomas ; Lang, Tobias ; Rehberg, Ingo ; Richter, Reinhard:
Spherical sample holders to improve the susceptibility measurement of superparamagnetic materials.
In: Review of Scientific Instruments. Bd. 83 (2012) Heft 4 . - 045106.
ISSN 1089-7623
DOI: https://doi.org/10.1063/1.3700185

Volltext

Link zum Volltext (externe URL): Volltext

Abstract

The design of two custom sample holders with a spherical cavity for commercial vibrating sample magnetometer systems is described. For such cavities, the magnetization M[over ->] and the internal magnetic field H(i)[over ->] of a sample are both homogeneous. Consequently, the material parameter M(H(i)) of a sample can be determined even for liquids and powders with a high magnetic susceptibility.

Weitere Angaben

Publikationsform: Artikel in einer Zeitschrift
Begutachteter Beitrag: Ja
Institutionen der Universität: Fakultäten
Fakultäten > Fakultät für Mathematik, Physik und Informatik
Fakultäten > Fakultät für Mathematik, Physik und Informatik > Physikalisches Institut
Fakultäten > Fakultät für Mathematik, Physik und Informatik > Physikalisches Institut > Ehemalige ProfessorInnen
Fakultäten > Fakultät für Mathematik, Physik und Informatik > Physikalisches Institut > Lehrstuhl Experimentalphysik V
Fakultäten > Fakultät für Mathematik, Physik und Informatik > Physikalisches Institut > Ehemalige ProfessorInnen > Lehrstuhl Experimentalphysik V - Univ.-Prof. Dr. Ingo Rehberg
Profilfelder
Profilfelder > Advanced Fields
Profilfelder > Advanced Fields > Nichtlineare Dynamik
Titel an der UBT entstanden: Ja
Themengebiete aus DDC: 500 Naturwissenschaften und Mathematik > 530 Physik
Eingestellt am: 09 Aug 2018 07:35
Letzte Änderung: 21 Nov 2023 14:12
URI: https://eref.uni-bayreuth.de/id/eprint/42926