Title data
Friedrich, Thomas ; Lang, Tobias ; Rehberg, Ingo ; Richter, Reinhard:
Spherical sample holders to improve the susceptibility measurement of superparamagnetic materials.
In: Review of Scientific Instruments.
Vol. 83
(2012)
Issue 4
.
- 045106.
ISSN 1089-7623
DOI: https://doi.org/10.1063/1.3700185
Abstract in another language
The design of two custom sample holders with a spherical cavity for commercial vibrating sample magnetometer systems is described. For such cavities, the magnetization M[over ->] and the internal magnetic field H(i)[over ->] of a sample are both homogeneous. Consequently, the material parameter M(H(i)) of a sample can be determined even for liquids and powders with a high magnetic susceptibility.
Further data
Item Type: | Article in a journal |
---|---|
Refereed: | Yes |
Institutions of the University: | Faculties Faculties > Faculty of Mathematics, Physics und Computer Science Faculties > Faculty of Mathematics, Physics und Computer Science > Department of Physics Faculties > Faculty of Mathematics, Physics und Computer Science > Department of Physics > Former Professors Faculties > Faculty of Mathematics, Physics und Computer Science > Department of Physics > Chair Experimental Physics V Faculties > Faculty of Mathematics, Physics und Computer Science > Department of Physics > Former Professors > Chair Experimental Physics V - Univ.-Prof. Dr. Ingo Rehberg Profile Fields Profile Fields > Advanced Fields Profile Fields > Advanced Fields > Nonlinear Dynamics |
Result of work at the UBT: | Yes |
DDC Subjects: | 500 Science > 530 Physics |
Date Deposited: | 09 Aug 2018 07:35 |
Last Modified: | 21 Nov 2023 14:12 |